Digital thermostat W1209 - description and review

grala1 
Hello everyone once again.
Today I will try to describe to you a very cheap Chinese digital thermostat W1209.





To buy this thermostat, you need to spend about $ 1.30 on Aliexpress, including shipping. Prices on Polish auction portals with shipping start from about PLN 14.
The device is sold complete with a temperature sensor.



Sometimes, when buying a thermostat, we have a choice of the color of the display, you can choose from: green, red, blue.
We also have the option of purchasing additional elements for this device - we can purchase a transparent colorless housing for approx. $0.8 with shipping or, for example, if we decide that our sensor has a too short cable (the sensor cable length is 0.5 m in the set) or it will be if it is damaged, we have a choice of sensors with wires of the following lengths: 0.5m, 1m, 2m.

parameters:
- supply voltage: 12VDC;
- minimum current consumption approx. 18mA, maximum approx. 66mA;
- temperature control range: -50 to 110°C;
- measurement resolution; 0.1°C;
- adjustable hysteresis: 0.1-30°C;
- temperature display resolution 0.1°C in the range -9.9°C - 99.9°C, the remaining range 1°C;
- resolution of temperature setting 0.1°C in the range -9.9°C - 99.9°C, in the remaining range the resolution is the same. but the display does not show decimal values;
- temperature measurement refresh rate: 0.5s;
- load capacity of the relay contacts: depending on the soldered relay - for me it is 20A at 14VDC and 20A at 125VAC;
- dimensions: 40x49x16mm - dimensions may vary depending on the relay used.

Our tested thermostat is built on the STM8S003F3 microcontroller. In addition, on the PCB we can find 3 buttons described as "SET", "+" and "-" for changing settings, 5V stabilizer, 3-digit, seven-segment display, temperature sensor socket, LED for visualization of relay activation, relay and 4-pin screw connector, to which we supply 12VDC (+12V, GND) and where we have the working contact of the relay (K0, K1).
In the power supply circuit, we have a diode D1 working there as a protection against reverse connection of the power supply.

After supplying the power supply, the display shows the current temperature from the sensor (as the sensor here the NTC thermistor 10k, 0.5%). If there is no temperature reading from the sensor, the display will show "888".
Changes of the temperature setting to which our thermostat will strive are made by briefly pressing the "SET" button. From now on, the displayed value starts flashing and you can change it using the "+" and "-" buttons. After making the change, we must press the "SET" button again, after which the display will switch to displaying the current temperature.
By holding down the "SET" button for 3 seconds, we enter the menu, where we have access to 8 different parameters:
- P0 - change of operating modes: heating (H) or cooling (C) - factory setting - C;
- P1 - hysteresis setting: 0.1 - 30°C - factory setting: 2.0;
- P2 - upper setting limit: -50 - 110°C - factory setting: 110;
- P3 - lower setting limit: -50 - 110°C - factory setting: -50;
- P4 - temperature measurement correction: -7 - 7°C - factory setting: 0;
- P5 - relay activation delay: 0 - 600sec - factory setting: 0;
- P6 - ??????: ON - OFF; factory setting: OFF;
- P7 - alarm after exceeding the temperature: OFF - 0 - 110°C - factory setting: 50;
- P8 - parameter reset and settings: C - H - factory setting: C

I failed to notice any changes after changing parameter P6. The descriptions that the sellers post do not match my version of the thermostat, even though both my thermostat and the seller's descriptions have the same name. You may also encounter a different version of the software.
Setting the C value in parameter P8 brings factory settings everywhere. The same can be done by supplying power to the device while pressing the "+" and "-" buttons.
When moving through the menu, if we want to change an option, we can press the "+" or "-" buttons several times, or if we want to significantly change the value, and we do not want to click dozens of times, we can hold one of these buttons longer, which will automatically change this value.
Exit the menu by pressing the "SET" button for a long time or it is enough to leave the selected parameter for 10 seconds and then the device will automatically exit the menu returning to displaying the temperature reading (the device will automatically exit the menu only if the parameter is displayed, e.g. P6, and not its value).
The thermostat remembers the settings after disconnecting it from the power supply. If, for example, we change the temperature setting, enter the menu, change the P1 and P2 parameters, exit the menu, enter the menu again and change the P4 parameter, but do not exit the menu and disconnect the power supply, the P4 parameter will not be saved, and the setting and parameters P1 and P2 will be saved.

Changing the operating mode. If we set parameter P0 to C, setting to 30°C, P1 to 2°C, then the relay will be activated when the temperature on the sensor is 32°C and it will be deactivated when the temperature is 28°C.
If we set parameter P0 to H, setting to 30°C, P1 to 2°C, the relay will be switched on when the temperature on the sensor is 28°C and it will be switched off when the temperature is 30°C.

The resolution of all settings is accurate to 0.1°C, but the device behaves as if the program was written on a 4-digit display - i.e. in the setting range -9.9°C to 99.9°C, one press of the button changes the setting by 0, 1 but outside this range, if we want to change e.g. the value 105 to 106, we must press the button as many as 10 times, because one pressing changes the setting by 0.1.

If the measured temperature exceeds the value declared in parameter P7, "---" will appear on the display.

In the absence of a reading from the measuring sensor, the state of the relay will remain unchanged - i.e. if, for example, we disconnect the sensor when the relay was on, the relay will remain in this state until we reconnect the sensor.

On the display we can still meet such characters as "LLL" and "HHH". This can happen if the temperature on the sensor reaches -50°C or over 110°C, or if we change the limits in parameters P2 and P3 and the current temperature is outside these settings.

Below is a diagram of the described thermostat.


The thermostat is quite cheap and as you can see it has some setting options.
We can use it, for example, to maintain the temperature of the solution when etching PCBs or, for example, in aquariums.

INin the next description I will try to present the W1219 thermostat with two seven-segment displays.
Here is the link to the article.
https://www.elektroda.pl/rtvforum/topic3437967.html#17076082

Update
As earlier on the forum I added a description of another thermostat - XH-W1219, I decided to compare the accuracy of temperature measurement between these versions of thermostats. As a reference, I used a thermometer built on the AVR processor and the Dallas DS18B20 sensor.
Here are some measurements:

DS18B20 W1209 W1219
87.2 93.1 87.0
85.0 90.8 84.7
82.3 87.2 82.3
75.7 79.9 75.7
65.0 67.7 65.0
56.0 57.9 56.1
49.6 50.8 49.7
40.8 41.5 40.8
28.3 28.5 28.4
17.1 17.1 17.0


As you can see, the W1209 thermostat begins to cheat us at higher temperatures - it overestimates the measurements and we can't change the P4 parameter here because after changing it, the measurement of lower temperatures will start to fail.
I thought that these thermostats are Chinese toys for a few zlotys, so maybe it's the fault of the thermistor used as a temperature sensor - some of its non-linearity or inaccuracy. Since both thermostats use exactly the same type of thermistor as a sensor, I decided to swap the measuring probes between the thermostats and measure again to see if it was the thermistor's fault.
Measurement results after replacing the measuring probes:

DS18B20 W1209 W1219
84.3 90,0 84,5
75,6 79,9 75,5
63,8 66,6 63,8
51,3 52,8 51,3
45,6 46,7 45,7
40,4 41,2 40,4
31,8 32,2 31,9
19,8 19,7 19,7
16,9 16.9 16.9


After taking these measurements, you can see that it is not the thermistor's fault, but the conversion of measurements into temperature by the software in the W1209 thermostat.

About Author
grala1 wrote 9755 posts with rating 4843 , helped 1476 times. Live in city Kalisz. Been with us since 2006 year.

Comments

Add a comment
REMUR. 24 Feb 2018 21:49

- P6 - alarm function, causes that after reaching the alarm temperature previously set in the thermostat, the function - P7 , the display will show "---", the relay will be disconnected and the... [Read more]

lelum0 24 Feb 2018 22:07

Just yesterday I ordered this system out of curiosity, because I did not believe that for this price you can get something sensibly working. I will need a differential thermostat in the future, has anyone... [Read more]

prosiak_wej 25 Feb 2018 00:02

I ordered a thermostat :) By the way, I grabbed such a panel voltmeter: https://obrazki.elektroda.pl/3997882000_1519513268_thumb.jpg ALIEXPRESS [Read more]

liroy123 25 Feb 2018 08:30

Hello, this thermostat module connected to me for a "moment" has worked for over 2 years, connected 24 hours a day to the network non-stop. He controlled the operation of the central heating pump and I... [Read more]

adrian18 25 Feb 2018 09:20

I would need a thermostat that would turn off the device to 0.1 degrees, as in the 70.2 Act, it should turn off at this temperature, not 0.1 earlier or 0.3 later. I don't know if this thermostat is... [Read more]

andrzej55 25 Feb 2018 09:45

Buddy adrian18, for more accurate control, we use PID controllers with a better quality sensor, e.g. Pt100. [Read more]

Strumien swiadomosci swia 25 Feb 2018 10:36

I assume it as a controller for liquid heaters in cars. Today it indicated -18 and since the previous winter it has been working so well. The cable is enough for the heater tubes and you don't need... [Read more]

HUGO 25 Feb 2018 12:00

Hello It's nice that there are such reviews, useful information for future buyers, keep it up :-) Regards Andrew [Read more]

wodzu12240 25 Feb 2018 12:39

Buddy Strumien Awareness, can you describe something more about these liquid heaters? Because I think I'm not the only one who would be interested? Thank you in advance :) [Read more]

Strumien swiadomosci swia 25 Feb 2018 12:45

Ordinary 3 or 4 glow plugs in a housing, from Focus, Peugeot Citroen Renault HDI, Volvo. You buy it for this 2 relays and control from this thermostat. I install it for faster heating of LPG. Installation... [Read more]

siewcu 25 Feb 2018 13:58

Do you think it would be possible to use it to measure the water temperature together with the heater control? You wrote about a load of up to 125 V, will the relay and the design of the device allow the... [Read more]

grala1 25 Feb 2018 14:14

The transistor that controls this relay on my board is S8050, we can pull up to 500mA on it. The relay that I have takes about 50mA. There is nothing to prevent you from putting a larger relay there,... [Read more]

vodiczka 26 Feb 2018 19:10

Suitable according to the description. Based on its price, I doubt it. :) [Read more]

prosiak_wej 26 Feb 2018 23:43

Out of curiosity, what kind of accuracy do you need? After all, even quartzes are not that sensitive (probably?). [Read more]

Anonymous 27 Feb 2018 11:27

That means, as usual, super "good" documentation/instructions. [Read more]

Jawi_P 28 Feb 2018 09:54

You are more likely to fail. Because you can buy, but I'm curious how you want to verify what you bought :) Contrary to appearances, temperature measurement in a wide range and accurate measurement... [Read more]

helazbyś 04 Mar 2018 11:24

My friend grala1 , you write "After powering up, the display shows us the current temperature from the sensor (as a sensor here the NTC thermistor 10k, 0.5%). If there is no temperature reading from... [Read more]

REMUR. 04 Mar 2018 11:45

@helazby the thermostat displays -0.1÷-9.9, i.e. with an accuracy of 0.1°C, while below it displays -10÷-50, i.e. with an accuracy of 1°C. Without a doubt, it can be used to measure the outside tem... [Read more]

robspeed 04 Mar 2018 23:03

Hello, my W1209 thermostat did not work at the beginning, the micro switch SET and the relay was not loaded, it was heating up ... Even under warranty it wasn't worth sending it back... Fortunately,... [Read more]